Fabrication Process for Ultra-Reliable Polyimide-Based Neural-Interface Technology

TitleFabrication Process for Ultra-Reliable Polyimide-Based Neural-Interface Technology
Publication TypeConference Paper
Year of Publication2025
AuthorsFluker, Jr., K. A., and J. W. Judy
Conference Name2025 23rd International Conference on Solid-State Sensors, Actuators and Microsystems (Transducers)2025 23rd International Conference on Solid-State Sensors, Actuators and Microsystems (Transducers)
Date Published08/2025
PublisherIEEE
Conference LocationOrlando, FL, USA
KeywordsImplant, Neural Interface, polyimide, Reliability
Abstract

Although polyimide can be used to produce high-channel-count neural interfaces with microelectrodes, their chronic reliability as an implant is a concern. In this paper we compared two different surface-treatment methods (O2 plasma and KOH/HCl soaks) to increase the strength and reliability of a bond formed between two fully cured layers of polyimide. Peel tests were used to quantify adhesion strength and were performed after test samples were soaked in RAA for days. The surface treatment based on O2 plasma formed a relatively strong bond (280 N/m) that was rapidly compromised by a 7-day RAA soak test. The bond formed by the surface treatment based on KOH and HCl soaks was far stronger (~500 N/m) and was not compromised by even a 15-day RAA soak test.

URLhttps://ieeexplore.ieee.org/document/11110719/http://xplorestaging.ieee.org/ielx8/11108179/11109079/11110719.pdf?arnumber=11110719
DOI10.1109/Transducers61432.2025.11110719